EasyManuals Logo

Apera Instruments 800 Series User Manual

Apera Instruments 800 Series
32 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #11 background imageLoading...
Page #11 background image
- 11 -
cup of distilled or deionized water, and rinse the electrode in it for a few seconds. Gently shake the
probe to remove excess water, and dry it with clean tissue paper (DO NOT rub or wipe the probe,
just use paper to dap off excess water). Gently stir the electrode for a few seconds after it’s dipped
into the test solution and then let it stand still. Wait for the stable measurement (a smiley face appears
and stays) and then record the readings. When test is finished, place the electrode in the storage
bottle and tighten the cap in order to keep the sensitivity of the pH sensor.
Notes:
1) This default 201T-F electrode will NOT give accurate and stable pH readings when testing purified
water such as distilled or deionized water. To measure distilled or deionized waters pH, users need to
use a specialized pH electrode with the meter such as LabSen
TM
803 Pure Water pH Electrode. Contact
us at 1-614-285-3080 or info@aperainst.com for more details. When testing purified water like spring
water or drinking water, it will take longer for the readings to get stabilized (typically 3-5 minutes) because
there is very few ions left to be detected by the sensor in those purified water.
2) Purified water such as distilled water and deionized water are recommended for only rinsing the
probe for the best result.
3) The 201T-F pH Electrode is NOT suitable for testing high-temperature samples (>176˚F). Testing
high-temperature samples could cause permanent damage to the electrode. A specialized electrode such
as LabSen
TM
213 (up to 225˚F) should be used in this situation.
4) The electrode housing is polycarbonate. When using cleaning solutions, take cautions on carbon
tetrachloride, trichlorethylene, tetrahydrofuran, acetone, etc., which will dissolve the housing and
invalidate the electrode.
5.3 Information regarding pH Calibration
5.3.1 Standard Buffer Solutions
The instrument adopts two series of standard buffer solutions: USA and NIST as shown in Table-4.
Users can select which one to use in P1.1 (refer to section 7.3)
Table-4 pH Standard Buffer Series
Calibration Icons
pH Standard Buffer Series
USA Series
NIST Series
3 Points
Calibration
1.68 pH & 4.00 pH
1.68 pH & 4.01pH
7.00 pH
6.86 pH
10.01pH & 12.45 pH
9.18 pH & 12.45 pH
5.3.2 pH Calibration Modes
Users can adopt 1 to 3 points to calibrate as needed. In the 3-point calibration mode, the 1
st
point
must be 7.00 pH (or 6.86 if using NIST). Then choose other calibration solutions to conduct 2
nd
and

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Apera Instruments 800 Series and is the answer not in the manual?

Apera Instruments 800 Series Specifications

General IconGeneral
BrandApera Instruments
Model800 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals