EasyManuals Logo

Apera Instruments 800 Series User Manual

Apera Instruments 800 Series
32 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #15 background imageLoading...
Page #15 background image
- 15 -
Press when reading is not stable during
calibration.
Press key when icon appears and
stays.
During calibration, readings being unstable for
over 3 minutes
1. Check whether there are bubbles in
glass bulb.
2. The pH Electrode is aged. Replace a
new one.
pH electrode zero electric potential out of
range (<-60mV or >60mV)
1. Check whether there are bubbles in
glass bulb.
2. Check whether pH buffer solution is
correct.
3. Check whether pH electrode is
damaged.
4. The pH electrode is aged. Replace a
new one
pH electrode slope out of range
(<85% or >110%)
The pre-set calibration due date is reached.
Time to do perform a new calibration
Press key to perform calibration or
cancel due calibration setup in parameter P1.2.
* If you find any air bubble in the glass bulb of the pH sensor, simply shake the probe for a few
times to remove it. The existence of an air bubble in the glass bulb will significantly decrease the
stableness of measurement.
5.5.3 The instrument has a function to return to factory default setting, which can be set up in P1.4
(refer to section 7.3). Returning to factory default setting is to restore the meter to theoretical value
(zero potential pH is 7.00, slope is 100%), and set all the parameters to default settings (see
appendix 1). When the meter’s calibration or measurement is performing abnormally, users can use
this function to let the meter return to factory default mode, and then conduct calibration and test
again. Please note that this function is irreversible once used.
5.6 Maintenance of the pH Electrode
5.6.1 Daily maintenance
The soaking solution contained in the supplied protective bottle is used to maintain activation in the
glass bulb and junction. Loosen the capsule, remove the electrode and rinse in distilled water
before taking a measurement. Insert the electrode and tighten the capsule after measurements to
prevent the solution from leaking. If the soaking solution is turbid or moldy, replace the KCL
solution (SKU: AI1107). Note: using other brands’ storage solutions could cause permanent
damage to the pH electrode because different chemicals are used. The electrode should not be
soaked in distilled water, protein solution or acid fluoride solution for long periods of time. In
addition, do not soak the electrode in organic lipids.
5.6.2 Calibration buffer solution
For calibration accuracy, the pH of the standard buffer solution must be reliable. The buffer solution

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Apera Instruments 800 Series and is the answer not in the manual?

Apera Instruments 800 Series Specifications

General IconGeneral
BrandApera Instruments
Model800 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals