- 19 -
7.3.3 Single point and multi-point calibration
If 1-point calibration is conducted after 3-point or 4-point calibration being done, the previous
calibration values in the same range will be replaced. In the meanwhile, the meter will display the
1-point calibration’s icon, and the other calibration icons will be removed, but the chip will still store
the data from the last calibration. When conducting multi-point calibration, users should follow the
sequence from low conductivity to high in case the high concentration solution contaminating the
low ones.
7.3.4 Reference Temperature
The factory default setting for reference temperature is 25. The reference temperature can be set
from 15˚C to 30˚C. Users can set it up in parameter setting P2.4 (see Section 7.4 for details).
7.3.5 Temperature compensation coefficient
The temperature compensation coefficient of the meter setting is 2.0%/°C. However, the
conductivity temperature coefficient is different from solutions and concentration. Please refer to
Table – 11 and the data collected during testing. Do the setting in P2.5. (see section 7.4 for more).
Note: When the coefficient for the temperature compensation is set to 0.00 (no compensation), the
measurement value will be based on the current temperature.
Table-11 Temperature compensation coefficient of special solutions
Solution
Temperature compensation coefficient
NaCl solution
2.12%/°C
5% NaOH solution
1.72%/°C
Dilute ammonia solution
1.88%/°C
10% hydrochloric acid solution
1.32%/°C
5% sulfuric acid solution
0.96%/°C
7.3.6 Precaution for calibration solution’s contamination
Conductivity standard solution has no buffer. Please pay attention not to contaminate the solutions
during usage (especially for standard solution of low concentration like 84μS/cm). The contaminated
standard solution will affect accuracy. Before submerging the electrode in standard solution, please
rinse the electrode with distilled water and allow it to dry.
7.3.7 Calibration Reminder Setup
Preset the interval between calibrations (starting from the time when
you set it), and then the meter will remind you to calibrate at the end of
that interval. For details, please see P2.2 (Section 7.4). When the preset
time is reached, Er6 icon will be displayed at the lower right corner of
Figure-11

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Apera Instruments 800 Series and is the answer not in the manual?

Apera Instruments 800 Series Specifications

General IconGeneral
BrandApera Instruments
Model800 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals