Interface:
Test Steps:
Step 1. Plug in the adopter than press the button and switch to the setting pressure mode.
Step 2. Long press and buttons and continue to press the button and its display will start
to blink.
Step 3. Adjust the pressure value by pressing the and buttons until the pressure value on the
pressure test point 4.0 cmH2O and press button.
Note: Remember to set the “Ramp Time “at 0 to save your testing time.
Step 4. Press the button to the standby mode, and then press button again to begin with air
pressure test to the first pressure test point 4.0 cmH2O, which the stable should be around 80
seconds. For iX Sense, the Pressure Accuracy should be in ± 0.5 cmH2O; for iX Fit, the Pressure
Accuracy should be in ± 1. Please use the same order to check the air pressure test point 10.0
cmH2O and air pressure test point 20.0cmH2O.