Questions and Answers

  • L
    Laura NavarroSep 18, 2025
    What to do if my Apex Digital iX Control Unit is showing Error E101-07?
    • C
      cindyodonnellSep 18, 2025
      If the device displays error E101-07, check for any damage in cable connections related to the heating humidifier. If the issue persists, replace the heating plate. If it still remains, replace the heating humidifier mainboard.
  • B
    blairwilliamAug 25, 2025
    Why is there no air flow through the mask of my Apex Digital iX?
    • T
      Tony JohnsonAug 25, 2025
      If there's no air flow through the mask, first, check that the power cord is properly connected and the device is turned on. Next, ensure the heated tubing is correctly connected. Finally, check for and unblock any obstructions in the air tubing.
  • D
    Daniel StephensAug 25, 2025
    Why is my Apex Digital iX showing 'Low P icon'?
    • J
      Joshua HoffmanAug 26, 2025
      The 'Low P icon' may appear if the mask isn't properly fitted. Make sure the mask fits securely on your face. Also, check the tubing for any damage and ensure it's properly connected. Inspect the pressure and flow sensor tubes to confirm they are correctly connected.
  • S
    Stacy ChambersAug 26, 2025
    What to do if my Apex Digital iX displays no display or incorrect display?
    • S
      Steven DanielAug 27, 2025
      If there is no display or an incorrect display, check if the power cord is connected to the power socket. If it is, the issue may be due to LCD or controlled PCB failure, in this case replace the LCD. If the problem persists, replace the mainboard.
  • G
    Gail HopkinsSep 19, 2025
    What to do if my Apex Digital iX Control Unit is showing Error E101-08?
    • D
      Daniel SanchezSep 22, 2025
      If error E101-08 appears, reconnect the heated humidifier. If the problem persists, replace the heating plate. If the issue still remains, replace the humidifier mainboard.
  • R
    ruizcarlaSep 23, 2025
    Why date cannot be copied to the SD card of my Apex Digital iX Control Unit?
    • G
      gthompsonSep 23, 2025
      If data cannot be copied to the SD card, check if the SD card is provided by APEX or check the SD card format (FAT32). Also, verify if the SD card is full and clear space for new data. If the issue persists, the data on the SD card may be corrupted; format the card or use a new SD card.
  • F
    felicia39Sep 9, 2025
    What to do if my Apex Digital iX is showing Error E101-01?
    • M
      Matthew RogersSep 9, 2025
      If error E101-01 is displayed, first reset the device. If the problem remains, the mainboard may need to be replaced.
  • P
    plopezSep 12, 2025
    What to do if my Apex Digital iX Control Unit is showing Error E100-002?
    • J
      Jon MillerSep 12, 2025
      If the device shows error code E100-002, restart it. After restarting, enter the Setting mode and select About to verify if the software version is displayed. If the problem persists, the 4G module may need to be replaced.
  • S
    Steven CombsSep 3, 2025
    What to do if my Apex Digital iX is showing Error 002?
    • J
      Jason ComptonSep 3, 2025
      If Error 002 is displayed, verify the SD card format is FAT32. Also, ensure the SD card is properly inserted into the slot and has sufficient capacity. If the problem persists, consider replacing the SD module.
  • N
    Nicholas AndersonSep 1, 2025
    What to do if my Apex Digital iX Control Unit is showing Error 006?
    • C
      Caitlin SmithSep 1, 2025
      If your device displays Error 006, ensure the cellular setting is turned on. If the signal remains weak, try replacing the 4G SIM card. Also, verify that the antenna connects to the 4G module board. If the problem persists, replace the 4G module board.

Summary

CHAPTER 1: Introduction & Intended Use

CHAPTER 2: Warning & Cautions

NOTE, CAUTION, AND WARNING STATEMENTS

Explanation of special statements for operating and maintenance procedures.

CHAPTER 3: Theory of Operation

CHAPTER 4: Maintenance

4-1. Cleaning of the Device

Procedure for cleaning the device exterior and internal circuit parts.

4-2. Cleaning the Humidifier

Instructions for daily cleaning of the humidifier and its water chamber.

4-3. Cleaning the Tubing

Guidelines for cleaning the air tubing, including checking for wear.

4-4. Replace Air Filter

CHAPTER 5: Troubleshooting

CHAPTER 6: Component Removal & Installation

6-1. Open the Top cover

Procedure to separate the top cover from the device by removing bottom screws.

6-2. LCD Display

Steps for removing and replacing the LCD display module on the iX Auto model.

6-3. Mainboard (iX Fit)

Instructions for removing and installing the mainboard for the iX Fit model.

6-4. 4 G & SD Module

Procedure for removing and installing the 4G and SD module board.

6-5. Blower & Air path module (iX Fit)

Steps for disassembling and removing the blower and air path module on iX Fit.

6-6. WIFI Board

Steps for removing and installing the Wi-Fi board on iX Sense and iX Auto models.

6-7. DC Jack cable

Procedure for disassembling and removing the DC jack cable.

6-8. Heated Humidifier

Steps for disassembling and installing the PCBA mainboard of the heated humidifier.

Chapter 7. Performance Verification

7-1. iX Fit; Sense Pressure Calibration

Procedure to calibrate pressure after replacing electronic components on iX Fit/Sense.

7-2 iX Fit; Sense Pressure Test

Steps to perform a pressure test to verify calibration on iX Fit/Sense models.

7-3 iX Auto Pressure Calibration

7-4 iX Auto Pressure Test

7-5 iX Series functional verification

Heated humidifier commutation functions

SD card function

Buzzer function

Chapter 8. Reset Total Meter

Chapter 9. Reset to Default Settings

Chapter 10. Service BOM

Chapter 10. Service Kit - Tools

Apex Digital iX Series Specifications

General IconGeneral
BrandApex Digital
ModeliX Series
CategoryControl Unit
LanguageEnglish