PG7 Rev. A2, 07/16
14
Automation Products Group, Inc.
Sample Rate (SAMPLE):
Adjusts the rate at which the gauge takes sample readings.
NOTE: Setting the Sample Rate to “SLOW” will help preserve battery life
(when applicable) and will also help to smooth rapidly uctuating readings.
Options: SLOW (4x/second), MEdIUM (8x/second), FAST (16x/second)
Setting the Sample Rate feature:
Step 1: Press the On/O button to enter the main setup menu.
Step 2: Press (Z) to cycle through the options until AdVSET is displayed.
Step 3: Press (S) to access the Advanced Settings menu.
Step 4: Press (Z) to cycle through the options until SAMPLE is displayed.
Step 5: Press (S) to access the Sample Rate setting options.
Step 6: Press (Z) to cycle through the setting options until the desired
setting is displayed.
Step 7: Press (S) to apply the displayed setting and return to advanced
setup menu.
Step 8: To exit the advanced setup menu, press (Z) until EXIT is displayed
and press (S) to exit to the main setup menu.