AquaLab 12 TROUBLESHOOTING
1. On the measurement screen, toggle to the correct model us-
ing the up and down arrow keys. The moisture content value
updates to correspond with the model selected.
2. If the correct model is not available, the model may not be
loaded on the instrument.
(a) To determine which models are loaded on the instrument,
cycle to the menu tab, select Moisture Content and then
the loaded models appear.
3. If the correct model is not available, load the appropriate model
using AquaLink 4 Software. The AquaLab DUO can hold a to-
tal of 100 models at any one time. You may need to remove
a model using the Software or use the delete option in the
%Moisture Calibration menu before you can add a new one.
AquaLink 4 saves any model that you remove with the soft-
ware and allows you to reload again later.
12. DUO PROBLEM:
Moisture Content displayed is not correct.
SOLUTION:
1. Model selected may not be correct for the product being tested.
(a) Toggle through the available models to find a more appro-
priate model.
(b) If the model is correct but not giving correct moisture con-
tent values it may be necessary to generate a new model
for the product or update an existing model. For informa-
tion about generating a model, contact Decagon Devices
for updating a model, refer to Section 10: DUO Opera-
tion.
13. DUO PROBLEM:
Moisture content does not show up on the screen.
SOLUTION:
79