56
Guidance and manufacturer’s declaration – electromagnetic immunity
Immunity test
IEC 60601-1-2
test level
Compliance level
Electromagnetic environment -
guidance
Electrostatic
discharge (ESD)
EN 61000-4-2
±2kV, ±4kV,
±8kV, ±15kV
air
± 8kV contact
±2kV, ±4kV, ±8kV,
±15kV air
± 8kV contact
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material
the relative humidity level should
be at least 30%.
Conducted
disturbances
inducted by RF
fields
EN 61000-4-6
3V in 0,15
MHz to 80
MHz
3V in 0,15 MHz to
80 MHz
Portable and mobile RF
communications equipment
should be used no closer to any
part of the product, including
cables, than 1.0m, if the
transmitter’s output power rating
exceeds 1W
a
. Field strengths
from fixed RF transmitters, as
determined by an
electromagnetic site survey,
should be less than the
compliance level in each
frequency range
b
.
Interference may occur in the
vicinity of equipment marked with
this symbol:
6V in ISM and
amateur radio
bands
between 0,15
MHz and 80
MHz
80% AM at 1
kHz
6V in ISM and
amateur radio
bands between
0,15 MHz and 80
MHz
80% AM at 1 kHz
Radiated RF
electromagnetic
field
EN 61000-4-3
Professional
Healthcare
environment
3 V/m
Professional
Healthcare
environment
3 V/m
80 MHz to 2,7
GHz
80% AM at 1
kHz
80 MHz to 2,7 GHz
80% AM at 1 kHz
Proximity fields
from RF
wireless
communications
equipment
EN 61000-4-3
385 MHz - 27
V/m
450 MHz - 28
V/m
710, 745, 780
MHz - 9V/m
810, 870, 930
MHz - 28 V/m
1720, 1845,
1970, 2450
MHz – 28 V/m
5240,5500,
5785 MHz -
9V/m
385 MHz - 27 V/m
450 MHz - 28 V/m
710, 745, 780 MHz
- 9V/m
810, 870, 930 MHz
- 28 V/m
1720, 1845, 1970,
2450 MHz - 28
V/m
5240,5500, 5785
MHz - 9V/m
Electrical fast
transient/burst
EN 61000-4-4
±1kV SIP/SOP
ports
±2kV AC port
100kHz
repetition
frequency
±1kV SIP/SOP
ports
±2kV AC port
100kHz repetition
frequency
Mains power supply should be
that of a typical commercial or
hospital environment.