16
SECTION
B
TROUBLESHOOTING
Display Screen Messages
Please read this section rst. If you still have questions, call Customer Service at
800.818.8877, 24 hours a day, 7 days a week, USA and Canada only. If outside of the
USA or Canada, contact your local ARKRAY distributor. Contact your healthcare professional
if you cannot reach Customer Service. Do not send your meter to ARKRAY or your local
dealerwithoutrstcallingtogetapprovaltodoso.
HI and Lo Messages
Themeterdisplaysresultsbetween20–600mg/dL.
HI Message
“HI”appearswhenthebloodglucoselevelisgreaterthan600mg/dLand
indicates severe hyperglycemia (much higher than normal glucose levels).
If“HI”isdisplayedagainuponretesting,pleasecontactthepatient’s
healthcare provider immediately.
Lo Message
“Lo”appearswhenatestresultislessthan20mg/dLandindicatessevere
hypoglycemia (very low glucose levels).
If“Lo”isdisplayedagainuponretesting,pleasecontactthepatient’s
healthcare provider immediately.
NOTE
• If messages for hyperglycemia or hypoglycemia are displayed, even if the
patient does not have those conditions, please contact Customer Service at
800.818.8877, 24 hours a day, 7 days a week, USA and Canada.
Error Messages
Er1: A used test strip was inserted.
Repeat the test with a new test strip.
Er2: The blood or control solution sample was applied before the
symbol appeared.
Repeat the test with a new test strip and wait until the symbol
appears before applying the blood or control solution sample.
Er4: The blood sample has abnormally high viscosity or insufcient
volume. Repeat the test with a new test strip.
Er5: This error message may appear when the wrong blood glucose test
strip is used instead of an Assure Prism multi Blood Glucose Test Strip.
Repeat the test with an Assure Prism multi Blood Glucose Test Strip.
Er6: There is a problem with the meter.
Do not use the meter. Contact Customer Service at 800.818.8877, 24 hours a
day, 7 days a week, USA and Canada.