11
SECTION
B
If results continue to fall outside the range printed on the test strip vial or on the idividually
wrapped test strip box, the test strip and meter may not be working properly. Do not use your
system and contact Customer Service at 800.818.8877, 24 hours a day, 7 days a week, USA
and Canada.
Comparing the Control Solution Test Results
The test result of each control solution test should be within the range printed on the label
of the test strip vial or on the individually wrapped test strip box. Repeat the control solution
test if the test result falls outside of this range. Out-of-range results may occur due to the
following factors:
Situations Actions
When the control solution bottle was not shaken well.
When the meter, test strip or the control solution were exposed to high or low temperatures.
When the rst drop of the control solution was not discarded or the top of the cap was not wiped clean.
When the meter is not functioning properly.
Repeat the control solution test.
When the control solution is past the expiration date printed on the bottle.
When the control solution is past its discard date (the date the bottle was opened plus three [3] months).
When the control solution is contaminated.
Discard the used control solution
and repeat the test using a new
bottle of control solution.

Other manuals for Arkray Assure PRISM MULTI

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Arkray Assure PRISM MULTI and is the answer not in the manual?

Arkray Assure PRISM MULTI Specifications

General IconGeneral
BrandArkray
ModelAssure PRISM MULTI
CategoryBlood Glucose Meter
LanguageEnglish

Related product manuals