41
If the meter is being operated by a second person who is providing
testing assistance to the user, the meter and lancing device should be
cleaned and disinfected appropriately when testing is conducted by the
second person.
IMPORTANT:
Check the expiration date printed on the test strip bottle. DO NOT use
expired Test Strips.
When you open a bottle of Test Strips for the rst time, record the
open date” on the test strip bottle. Discard the bottle and any
remaining Test Strips 3 months after the open date.
After taking a Test Strip out of the bottle, close the cap of the bottle
immediately.
Use each Test Strip immediately after taking it out from the bottle.
Do not use wet, bent, scratched, or visibly damaged Test Strips.
Keep the Test Strips away from direct sunlight and heat. Store the
test strip bottle in a dry, cool place