Cleaning and Disinfecting Lancing Device
Clean and disinfect Lancing Device after every use with DISPATCH® hospital
cleaner disinfectant towels with bleach, with EPA registration number 56392-
8. Other wipes with EPA Reg. No. 56392-8 may also be used.
1. Inspect for blood, debris, dust, or lint anywhere on the lancing device.
2. Clean or disinfect lancing device.
3. To clean the lancing device and endcap, blood/body uids must be
thoroughly cleaned from surface and objects beore disinfecting.
Dispense the towel (DISPATCH) from its package. Wipe surface
with towel (DISPATCH) until completely wet. Allow to remain wet for
one minute at room temperature (68-77° F). Discard used towel in a
sealed container where it will not be touched by others.