Page 22 · 5400 Series TECSource User’s Manual
Menu Description
Factory
Default
Set While Rmt
To disable changing the set point while in
Remote Mode, set this value to No. To
enable changing the set point while in
Remote mode, set this Value to Yes.
No
Err While Rmt
To turn off the display of errors while in
remote mode, set this value to No. To
display errors while in remote mode, set
this value to Yes.
Yes
Terminal
Mode
Terminal mode simply echoes any
characters received over the serial or USB
interfaces.
No
Msg Term
This controls the output message
termination, and can be set to CR/LF, CR,
LF, or None.
CR/LF
System Menu
Config
Allows for restoring the instrument to
defaults, saving configurations, and
restoring configurations. See Saving and
Restoring Configurations below for more
information.
No Change
Brightness
The vacuum florescent display can be set
to one of eight brightness levels.
100%
Audible Beep
This setting controls when the unit
produces audible feedback. Set to No to
prevent sound, or Yes to enable audible
alerts such an error messages.
Yes
Lockout Knob
Lockout knob allows you to disable knob
operation from the main display. This
prevents accidental changes of the set
point. The knob will always work in the
menus regardless of this setting.
No
Knob Speed
Controls the adjustment speed of the knob.
Possible values are slow, medium, and fast.
Slow
Out Off Menu
Thi Limit
Turns off the TEC output if the TEC exceeds
the temperature high limit
Yes