Charging Rack Installation
INSTALLATION AND OPERATION MANUAL
CR3 CR5, and CR7 Charging Racks
2 Charging Rack Installation
2.1 General
The unit shall be installed by authorized personnel only.
The units shall be placed in a dry environment with a temperature range from +5° C up to + 40° C (41° F to
104° F).
The units shall be mounted on a vertical wall.
Avoid mounting the Charging Rack in a sunlit place. This can affect the charging capacity.
Avoid mounting the Charging Rack where radio/network coverage is not sufficient. This can reduce the
messaging capacity.
The unit can be mounted to a vertical wall (concrete or plaster walls only) no higher than 2000 mm or be
placed on table if table adapters are used.
If the Charging Rack is connected to a power supply via an AC wall plug, serial power supply connection
is not allowed.
If the Charging Rack is connected to a power supply via an AC wall plug, the socket-outlet shall be
situated near the equipment and shall be easily accessible.
If the Charging Rack is connected to a power supply via a fixed connection, a readily accessible
disconnect device shall be incorporated in the building installation wiring.
Delivery Includes
Charging Rack
Power supply cord including IEC C14 connector (male)
Inlet accessory kit including cable support holders and screws
Table adapters
An extension cord IEC C13 AC connector (female, IEC60320-C13) to wall socket has to be ordered
separately. It shall be connected between the pre-installed IEC C14 AC connector (male) and the
wall socket.
Required Tools etc.
Screwdrivers
Cutting pliers
Multimeter
Screws and wall plugs for wall mounting.
Make sure the screws and wall plugs have the correct length for the type of wall used.