ERRORS AND FAULTS
See the manuals CDROM for further information.
1. ERROR MESSAGES
The ATEQ F5200 can display error messages if there are operational problems.
Reference fault / Test fault / Pressure in excess of the full scale / Error on the
differential sensor / Pressure in excess of the max. threshold / Pressure below the
min. threshold / ATR fault / fault or drift CAL / Valve commutation fault / PROG Error /
Inappropriate size for the selected unit of pressure / Sealed component learning error /
Sealed component error / Sealed components large leak error / Auto-test fault /
2. IN CASE OF OPERATION DOUBT
If a test machine begins to detect too many fail parts (more than three consecutively), it
is advisable to carry out a check on the whole unit. The quality of the manufacture and
operation of the leak detector should be the last things considered.
There is a possibility that the seals may be cut by shavings or worn by repetitive
squashing. This can be prevented by regular servicing and replacement of the seals.
If all the other checks do not resolve the problem, the instrument’s circuit may be
checked.
ATEQ does not accept any liability in regard to calibrations and settings to its
instruments which are not carried out by its own personnel.