ON/OFF Control
1. With the concentration adjusted to the desired level, the
aerosol output can be switched on and off as needed by using
the Aerosol Switch. For visual reference the Aerosol Switch
will illuminate green when it is ON.
2. When switching off the aerosol, it will take approximately 20
seconds for the aerosol to stop. The inert gas will continue to
flow.
3. Turning ON the Aerosol Switch will restart aerosol production
at the previous setting.
Shut-down Sequence
Caution
Failure to follow these instructions will result in premature
equipment failure.
1. Turn the Aerosol Switch to OFF.
2. Close the metering valve by fully turning the knob clockwise.
3. Allow all visible aerosol to purge out of the output collar.
4. Turn off electrical power to the generator at the Main Power
ON/OFF Switch.
5. Allow inert gas to flow through the generator for a minimum of
one (1) minute before turning off the gas at the regulator. This
purges the system of residual aerosol or wet liquid aerosol
reagent.
Warning – High Pressure Gas Hazard
Always allow gas in the line to DECOMPRESS before
disconnecting the gas line at the generator.
6. Disconnect the gas line, power cord and any connections to
the output collar.
Warning – Burn Hazard
Severe burns may occur if in contact with the output collar.
7. Allow the generator to cool to a temperature below 80ºC
before transporting.