Note
The power cord can be reattached and the Main Power Switch
turned ON temporarily to verify the temperature.
8. If the generator will not be used for a more than 24 hours, the
reservoir should be drained of liquid aerosol reagent into a
properly labeled container.
Warning – Fire Hazard
DO NOT drain the liquid aerosol reagent until the generator
has completely cooled to ambient conditions.
Warning – Fire Hazard
DO NOT perform any maintenance until the generator has
completely cooled to ambient conditions.
Changing to Different Liquid Aerosol Reagents (Non DOP)
Note
Changing to a different liquid aerosol reagent applies to 5C Part
Numbers 9300207 and 9300208 only. This does not apply to
9300213 or 9300214 which are for DOP only.
Before changing to a different liquid aerosol reagent, perform the
following purging procedure.
1. Drain all previous liquid aerosol reagent from reservoir.
2. Replace drain plug and tighten.
3. Turn the unit ON and allow the unit to reach the operating
temperature.
4. Connect the insert gas line and turn on the compressed insert
gas
5. Turn the Aerosol Switch to ON.
6. Fully turn the metering valve knob counterclockwise.
7. Allow the previous liquid aerosol reagent left in the unit to
purge out until the there is only insert gas flowing.
8. Follow the Shut-down Sequence in this manual
9. The 5C is ready for use with a different liquid aerosol reagent.