Thermal Generator - 5C C-2
Improper operating temperature
Switch off the aerosol and wait
a minimum of 25 minutes for
temperature stabilization or
return the unit for service.
Improper inert gas pressure
Verify the inert gas pressure is
set to 50 psi ± 5.0 psi (3.45 Bar ±
0.35 Bar).
Inert gas line is blocked, kinked or
leaking
Check inert gas line. Replace
as necessary
Internal regulator is defective Return the unit for service
Internal filters are clogged Return the unit for service
Internal solenoid valve is defective Return the unit for service
Wet Aerosol
Generator was not purged before
starting operation
Turn off Aerosol Switch and
allow gas flow to purge for one
minute.
Improper inert gas pressure
Verify the inert gas pressure is
set to 50 psi ± 5.0 psi (3.45 Bar ±
0.35 Bar).
Improper operating temperature
Switch OFF the aerosol and
wait a minimum of 25 minutes
for temperature stabilization or
return the unit for service.
Internal filters are clogged Return the unit for service
Internal regulator is defective Return the unit for service
No Display on
Controller
Not plugged to appropriate voltage
source
Verify the unit is connected to
an output with the correct
voltage
Fuse or fuses are blown
Check fuses & replace if
necessary (See Section 6.2)
Main Power Switch in OFF position
Turn Main Power Switch to the
ON position
Temperature controller is defective Return the unit for service
from inside unit (not
only from output
nozzle)
Improper inert gas pressure
Verify the inert gas pressure is
set to 50 psi ± 5.0 psi (3.45 Bar ±
0.35 Bar).
There is gas leakage inside the unit Return the unit for service