EN Power Focus 600
38
© Atlas Copco Industrial Technique AB - 9836 4954 01
1. On the Tool menu, go to Cali-
bration, and then Configura-
tion.
2. Press OK to enter the Configu-
ration dialog.
3. Calculate the new calibration
value using the equation above.
• Make at least three tighten-
ings and use the torque val-
ues read from the controller
and the reference transducer
(ACTA or similar) for calcu-
lating the new calibration
value.
• The old calibration value is
shown in the Calibration
value box.
4. To store the new calibration
value in tool memory, enter the
value in the Calibration value
text box by using the keypad and
then press OK.
5. To activate the alarm for when
the next calibration is to be per-
formed, go to Show alarm and
select On.
6. To set the alarm interval, go to
Next calibration date and then
press OK.
• To keep the calibration inter-
val to once a year just press
OK, as the default value is
set to be the same date a year
from today provided you
have set the time and date
correctly in Settings->Pref-
erences. For more informa-
tion, see section Date and
time [Page000].
• To set another date enter it
on the format yyyymmdd
and then press OK.
7. To store the settings go to Apply
and press OK.
Figure: Tightening menu - Configuring Pset
Accessories setup
{ , : profile4 in element [iosection]}
A tutorial on how to connect and set up the controller with accessories such as a
Socket selector and Scanner.