26
5.1.5 Medicament sprayers
The sprayer tube must be exchanged after each patient.
Dismount the medicament sprayer (fig. 32, 506.5120.0;
g.33,506.5225.0)andthoroughlyrinseallpartsunderrun-
ning water. You may, of course, add a cleanser or detergent.
Use water to thoroughly rinse all residues of these
substances.
Make sure that the air opening is not closed!
 When xingthe twin tube nozzle again,makesure
that the mark (0,X or milling area) on the nozzle shows
upwards !
Sprayertubesareavailableasspareparts(seeg.32/
g.33).
000.0237.0
055.0029.0
000.0219.0
506.5121.0
005.0026.0
000.0577.0
Fig. 32.
506.5120.0
505.0280.0
055.0028.0
005.0026.0
000.0577.0
506.5227.0
(055.0028.0)
(505.0280.0)
(505.0258.0)
(005.0026.0)
(055.0021.0)
Fig. 33.
506.5225.0
5.2 Endoscope quivers
The metal quivers of the endoscope holder are to be
usedsolelyforholdingtheendoscopes,thesersthaving
been cleaned. The quivers are to be cleaned daily and
subsequently disinfected. For doing this, the stopper at
the lower end should be taken off.
5.1.6 Instrument trays
Before disinfection, thoroughly rinse the trays under running
water. You may, of course, add a cleanser (deter-gent) or
surface disinfectant.
Thoroughly rinse the trays under running water.
Melamine and anodized aluminium trays cannot be
sterilised.
5.3 Ear rinsing bowl
The ear irrigation bowl is not autoclavable!
Cleaning and disinfection (also machine cleaning) up to
93°C.
5.0 Cleaning
!
upwards
!
Sprayertubesareavailableasspareparts(seeg.32/
g.33).