12.0 Notes on EMC
12.1 Guidelines and Manufacturer´s Declaration - Emissions
The ATMOS
®
C31isintendedforuseintheelectromagneticenvironmentspeciedbelow.Thecustomeroruserof
the ATMOS
®
C 31 should ensure that it is used in such an environment.
Emissions Test Compliance Electromagnetic Environment - Guidance
RFEmissions
CISPR11
Group 1 The ATMOS
®
C31usesRFenergyonlyforits
internalfunction.Therefore,itsRFemissionsare
very low and are not likely to cause any interference
in nearby electronic equipment.
RFEmissions
CISPR11
Class B
The ATMOS
®
C 31 is suitable for use in all esta-
blishments, including domestic, and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Harmonics IEC 61000-3-2 Class B
Flicker
IEC 61000-3-3
is conform
12.2 Guidelines and Manufacturer´s Declaration - Immunity
The ATMOS
®
C31isintendedforuseintheelectromagneticenvironmentspeciedbelow.Thecustomeroruserof
the ATMOS
®
C 31 should ensure that it is used in such an environment.
Immunity Test
IEC 60601-
Test Level
Compliance Level
Electromagnetic
Environment - Guidance
ESD
IEC 61000-4-2
± 6 kV Contact
± 8 kV Air
± 6 kV Contact
± 8 kV Air
Floors should be wood, concrete,
orceramistile.Ifoorsaresynthe-
tic, the relative humidity should be
at least 30%.
EFT
IEC 61000-4-4
± 2 kV Mains
± 1 kV I/Os
± 2 kV Mains
Inapplicable
Mains power quality should be that
of a typical commercial or hospital
environment.
Surges
IEC 61000-4-5
± 1 kV Differential
± 1 kV Common
± 2 kV Differential
± 1 kV Common
Mains power quality should be that
of a typical commercial or hospital
environment.
Power Frequency
50/60 Hz
Magneticeld
IEC 61000-4-8
3 A/m Inapplicable Powerfrequencymagneticelds
should be that of a typical com-
mercial or hospital environment.
■MedicalelectricalequipmentissubjecttospecialprecautionswithregardtoEMCandmustbeinstalledacc.
to following EMC notes.
■PortableandmobileHFcommunicationfacilitiescaninfluencemedicalelectricalequipment.
■Theuseofotheraccessories,otherconvertersandcablesthanstatedmayleadtoanincreasedemission
or a reduced interference immunity of the equipment or system.
The device may not be used directly next to other devices or piled up with other devices.
If operation next to or piled with other devices is necessary, please watch the device to check its intended
operation in this arrangement.