40
If, nevertheless, the errors cannot be removed inform the ATMOS service staff. Do not start any attempts to repair the unit
yourself !
Pay also attention to corresponding chapters in separate operating instructions !
Rinsingbasindoesnotproperly
empty
• clean or replace the sieve
• have waste water hoses checked by
a service technician or plumber
7.11 Kidney-shaped rinsing basin
– sieve in rinsing bowl is clogged
– waste water hoses are clogged
Failure with
the Unit
• decalcify rinsing head
• replace rinsing bowl
–rinsingheadisclogged(calcied)
– rinsing bowl is leaking
Basin rinsing does not function at
alloronlyinsufciently
7.0 Trouble-shooting
7.12 Endoscope support
Endoscope supports are no longer
heated
• Have the switch / control / sensor
exchanged by your service technician.
– overtemperature protection switch
was activated
– control is defective
– temperature sensor is defective
Malfunction at
the unit