EasyManua.ls Logo

Axon MultiClamp 700A - Page 131

Default Icon
159 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Reference Section 119
Chapter 5
However, it should be remembered that dynamic voltage errors and bandwidth errors
can still occur in the above example, because these depend on R
s
and C
m
and not on the
size of the membrane current. Even if you are measuring only small membrane
currents in a whole-cell recording, application of R
s
compensation can greatly improve
the fidelity of the voltage clamp.
As a general rule, it is best to try R
s
compensation to see if it makes a difference. This
is certainly advisable in all whole-cell recordings. Compensation is rarely useful with
isolated membrane patches, which typically have small capacitance and membrane
currents. Indeed, the Whole Cell controls (which must be set before using R
s
compensation) are disabled with the 5 and 50 G feedback resistors typically used for
isolated patch recordings. An exception is macropatches or nucleated outside-out
patches, in which the currents can be quite large and for which R
s
compensation may
be necessary.
If R
s
compensation is found not to be necessary, it is best to turn it off. This is because
R
s
compensation increases noise.
Adjusting R
s
Compensation
It is recommended that you practice adjusting R
s
compensation with the PATCH-1U
model cell before using compensation in a real experiment. (See Chapter 5,
MODEL
CELL
.) Connect the CELL connector to the CV-7A headstage. Set the feedback
resistor to 500 M (for Voltage Clamp) and Seal Test to 100 mV at 50 Hz.
When adjusting R
s
compensation, the Filter on Scaled Output should always be set at
10 kHz or above. Check the Seal Test checkbox and observe Membrane Current at a
fast sweep speed on an oscilloscope, triggering the oscilloscope so you can clearly see
the rising edge of the signal (Figure 4.15).

Table of Contents

Related product manuals