PROBLEM POSSIBLE CAUSES REMEDIES
Screen is blank 1. Battery is dead. 1. Replace battery.
Will not count in totalizer mode 1. Improperly programmed. 1. Check programming.
2. Broken or defective wiring. 2. Check wiring.
3. Improperly connected. 3. Check connections.
4. Transmitter defective. 4. Repair or replace transmitter.
Will not indicate rate of flow 1. Improperly programmed. 1. Check programming.
2. Improperly connected. 2. Check connections.
3. Transmitter defective. 3. Repair or replace transmitter.
Cannot program 1. Program enable jumper is not 1. Install jumper.
installed or installed improperly.
Cannot reset from front panel 1. Reset enable is not 1. Reprogram mode #6 to ON.
programmed.
Erroneous readings 1. Improperly programmed. 1. Check programming.
2. Defective transmitter. 2. Repair or replace transmitter.
No Pulse Output 1. Defective output Transistor 1. Replace ER-9
2. Improper wiring 2. Check connections
Field calibration consists of determining the exact transmitter pulse
output per unit of measure for your particular meter/transmitter com-
bination and then using this value as the transmitter pulse output value
when calculating the counter and time base values on page 4.
The procedure is as follows:
1. Set the totalizer scale factor to "1".
2. Set the totalizer Decimal Point to "off".
3. Reset the counter to zero.
4. Run fluid into a weigh tank or calibrated vessel.
5. Determine number of pulses per gallon by dividing indicator
reading by number of gallons of fluid in the vessel. Use this value
for your calculations.
Troubleshooting
Field Calibration
Example:
You programmed the indicator for calibration and connected the outlet
of a 1" OP meter to a calibrated vessel. You opened the valve and
allowed fluid to flow into the vessel. You determined that there was
22.35 gallons of fluid in the vessel. The reading on the indicator is 1720.
1720 / 22.35 = 76.95
The transmitter output is 76.95 pulses per gallon. Use 76.95 when
calculating the Totalizer Scale Factor on page 5.
6
For further assistance, call our Technical Support Staff at 414-355-0400, Extension 637.

Other manuals for Badger Meter ER-9

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Badger Meter ER-9 and is the answer not in the manual?

Badger Meter ER-9 Specifications

General IconGeneral
BrandBadger Meter
ModelER-9
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals