Configuration Accepted
Configuration Not
Accepted
Name Description
Reference Point Signal
Reference points have moved. The display sector in the direction of the
moved point is shown in blue.
Note: In the configuration not accepted state, the OSSDs will be off even if the display says GO (safety
zone clear). The configuration must be accepted before the OSSDs will turn on.
9.4 Diagnostic Notes, Warnings, and Errors
Icon Fault Code Device Status
OSSD
Status
Description
DLDNF Normal Off Downloading new firmware.
DLDNC Normal Off Downloading new configuration.
CLEANW2 Normal On Clean the window to avoid a lockout condition.
ITLOCKx Normal Off Interlock; waiting for a restart signal to turn the relevant OSSDs back on.
!
INTF6
INTF6 Normal On
Non-safety related internal test failure.
Reset the system using the reset function or cycle the power to the device.
Using a separate filtered power supply and/or a grounded shielded cable to
connect the scanner to the cabinet can eliminate external causes of the
failure.
If the failure persists, contact the factory for support.
BOOTF Normal Off
Invalid boot. Re-boot the system until the normal condition is restored. If
warning persists, contact the factory for support.
MUT
TIMEOUT
Normal On
Muting has expired because it is maintained beyond the maximum timeout
time.