Troubleshooting I 42
Accuracy
± 0.5% F.S.
Calibration Points
1 to 5 points
Calibration Solutions
10 µ S/cm, 84 µ S/cm, 1413 µ S/cm,
12.88 mS/cm, 111.8 mS/cm
Temperature Compensation
0 to 100° C (32 to 212° F)
Temperature Coefficient
Linear (0.0 to 10.0%/° C)
Pure water
Reference Temperature
20° C or 25° C
Cell Constant
K = 0.1, 1, 10 or custom
TDS
Range
0.00 to 10.00 ppt (max. 200 ppt)
Resolution
0.01, 0.1, 1
Accuracy
± 1% F.S.
TDS Factor
0.1 to 1.0 (default 0.5)
Salinity
Range
0.00 to 80.00 ppt, 0.00 to 42.00 psu
Resolution
0.01
Accuracy
± 1% F.S.
Resistivity
Range
0.00 to 20.00 MΩ
Resolution
0.01, 0.1
Accuracy
± 1% F.S.
Dissolved Oxygen
Range
0.00 to 20.00 mg/L
0.0 to 200.0% saturation
Resolution
0.01, 0.1
Accuracy
± 0.2 mg/L, ± 2.0%
Calibration Points
1 or 2 points
Temperature Compensation
0 to 50° C (32 to 122° F)
Pressure Correction
60.0 to 112.5 kPa, 450 to 850 mmHg
Salinity Correction
0.0 to 50.0 g/L
Temperature
Range
0 to 105° C (32 to 221° F)
Resolution
0.1° C (0.1° F)
Accuracy
± 0.5° C (± 0.9° F)
Calibration Point
1 point
Other Specifications
Memory
500 data sets
Communication Interface
USB-B
Operating Temperature
0 to 50° C (32 to 122° F)
Storage Temperature
0 to 60° C (32 to 140° F)
Relative Humidity
< 80% (non-condensing)
Display
LCD, 125 × 100 mm (4.9 × 3.9 in.)
Power Requirements
DC 5V/400mA power adapter
Dimensions
210 (L) × 188 (W) × 60 (H) mm
(8.2 × 7.4 × 2.3 in.)
Weight
1.5 kg (3.3 lb)
Troubleshooting
Fault
Cause and Corrective Action
Screen shows
Electrode dried out.
Soak the pH electrode in 3M KCl solution for about
30 minutes.
Soak the ion selective electrode in 100 ppm standard
solution for about 30 minutes.
Soak the conductivity electrode in tap water for
about 10 minutes.
Measurement exceeded the maximum range.
Check the electrode and sample.
Drifting erratic
readings
Check whether electrode is contaminated, clogged
or broken.
Screen shows
pH buffer problem. Use freshly prepared buffer
solutions to calibrate the meter.
Electrode has expired. Replace pH electrode.
DO electrolyte solution is depleted.
Refilling electrolyte solution.
Zero oxygen solution is contaminated.
Replace the calibration solution.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Bante Instruments 9 Series and is the answer not in the manual?

Bante Instruments 9 Series Specifications

General IconGeneral
BrandBante Instruments
Model9 Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals