29K5902142 /04 Nio 5MP
4.16.3 DICOM compliance check
About DICOM compliance check
The DICOM compliance check will measure the DICOM curve of your display in different steps. After
measurement, the DICOM status report is shown.
To start DICOM compliance check:
1. Bring up the OSD main menu.
2. Navigate to the Configuration > Calibration > Embedded QA menu.
3. Select DICOM Compliance Check to start the compliance check.
Warning: Pressing a key during the compliance check will abort the check.
4.16.4 DICOM calibration
About DICOM calibration
The DICOM calibration will add a correction to the current DICOM curve to approach the perfect DICOM curve
as well as possible.
To start DICOM calibration:
1. Bring up the OSD main menu.
2. Navigate to the Configuration > Calibration > Embedded QA menu.
3. Select DICOM calibration to start the calibration.
Warning: Pressing a key during calibration will abort the calibration, previous values will be
restored.
Note: After calibration, the compliance check will start automatically.
4.16.5 Reset DICOM calibration
About reset DICOM calibration
It is possible to restore the original (factory default) DICOM curve.
To reset the DICOM calibration:
1. Bring up the OSD main menu.
2. Navigate to the Configuration > Calibration > Embedded QA menu.
3. Enter the DICOM Preferences submenu.
4. Select Reset DICOM Calibration to restore the original (not corrected) DICOM curve.
4.16.6 DICOM error threshold
About DICOM error threshold
The threshold to define the DICOM compliance can be modified in steps of 5% starting from 5 to 30%. When
the maximum deviation is not bigger than the selected threshold, the compliance check will be OK.
To set the DICOM error threshold:
1. Bring up the OSD main menu.
2. Navigate to the Configuration > Calibration > Embedded QA menu.
Advanced operation