K5902133 (451920612483)-02 MDSC-8255
45
US and Canada: hospital grade cord-set has to be used, provided with instructions to indicate that
grounding reliability can be achieved only when the equipment is connected to an equivalent receptacle
marked hospital only or hospital grade. These instructions need to be marked either on the equipment or
on a tag on the power cord.
Transient over-voltage
If the device is not used for a long time, disconnect it from the AC inlet to avoid damage by transient over-
voltage.
To fully disengage the power to the device, please disconnect the power cord from the AC inlet.
Connections
Any external connection with other peripherals must follow the requirements of clause 16 of IEC60601-1 3rd
ed. or Table BBB.201 of IEC 60601-1-1 for the medical electrical systems.
Connection of PEMS by network/data coupling to other equipment (MDSC-8255 MNA
version only)
Connection of the PEMS to a network/data coupling that includes other equipment could result in
previously unidentified risks to patients, operators or third parties.
The responsible organization should identify, analyze, evaluate and control these risks.
Subsequent changes to the network/data coupling could introduce new risks and require additional
analysis.
Changes to the network/data coupling include:
- changes in network/data coupling configuration;
- connection of additional items to the network/data coupling configuration;
- disconnecting items from the network/data coupling configuration;
- update of equipment connected to the network/data coupling configuration;
- upgrade of equipment connected to the network/data coupling configuration
Water and moisture
The equipment is IP20 compliant. The monitor front side is IPX5 compliant (for versions with front protection
glass only).
Moisture condensation
Do not use monitor under rapid temperature and humidity change condition or avoid cold air from air-
conditioning outlet directly.
Moisture may condense on the surface or inside of the unit, or create a mist residue inside the protection
plate, this is not a malfunction of the product itself, although it may cause damage to the monitor.
If condensation happens, let the monitor stand unplugged until there is no condensation.
Ventilation
Do not cover or block any ventilation openings in the cover of the set. When installing the device in a cupboard
or another enclosed location, heed the necessary space between the set and the sides of the cupboard.
Installation
Place the equipment on a flat, solid and stable surface that can support the weight of at least 3 units. If you
use an unstable cart or stand, the equipment may fall, causing serious injury to a child or adult, and serious
damage to the equipment.
Do not allow to climb or rest on the equipment.
When adjusting the angle of the equipment, move it slowly so as to prevent the equipment from moving
from or slipping off its stand or arm.
When the equipment is attached to an arm, do not use the equipment as a handle or grip in order to move
the equipment. Please refer to the instruction manual of the arm for instructions on how to move the arm
with the equipment.
Provide full attention to safety during installation, periodic maintenance and examination of this equipment.
Important information