K5902133 (451920612483)-02 MDSC-8255
55
Contrast ratio 1400:1 (typical)
LCD response time (Tr + Tf) 8 ms (typical)
White point Native: 10000K (Typical)
Calibrated: 5600K, 6500K, 7600K, 9300K
Gamma Native, 1.8, 2.0, 2.2, 2.4, DICOM, Video
Housing color RAL 9003
Screen protection 2-side anti-reflective alkali-aluminosilicate glass (no screen protection on
NG versions)
Keyboard Membrane keyboard
Video input signals 4K-UHD input
• 1x DP 1.1 up to 3840 x 2160 @30Hz
• 2x DP 1.1 up to 1920 x 2160 @50Hz/60Hz
• 1x DP 1.2 MST up to 3840 x 2160 @50Hz/60Hz
2x FO SFP+ for 4K-UHD Nexxis link
FHD input (upscaled to UHD)
• 1x DVI
• 1x 3G-SDI
Video output signals 1x 3G-SDI (3G-SDI input loopthrough)
1x DVI (image on screen downscaled to FHD)
Video formats • Display Port 1.2 MST (10 bit) up to 3840x2160 @60Hz RGB 30bits/pixel
• Dual stream DP 1.1 (10-bit) up to 1920x2160 x2 @60Hz RGB 30bits/
pixel
• HDMI 2.0 up to 3840x2160 @60Hz RGB/YCbCr (4:2:0/4:2:2/4:4:4) with
HDCP 2.2 and 1.4
• HDMI 1.4 up to 3840x2160 @30Hz RGB/YCbCr (4:2:2/4:4:4) with HDCP
1.4
Integrated Nexxis 4k decoder MNA-240 decoder integrated
Remote control FW download, DDC protocol on DVI and DP auxiliary channel (on DP main
connector) & control protocol through network connection
Power consumption (Max) 170 W ± 10%
Power requirements 100-240 VAC / 50-60 Hz / 1.8-0.9 A
DC power output DVI connector: +5V on pin 14 & 16 / 500mA
DP connector: +3.3V / 500mA
USB connector: +5V / 1A
DC out connector: +5V / 2A
Power management Low power mode: 52 W (typical)
Power-off: ~ 1 W
Dimensions display (W x H x D) 1265 x 770 x 85 mm (49.8 x 30.3 x 3.3 in)
Net weight display With front protection glass: 33.6 kg (74.1 lbs)
Without front protection glass : 24.6 kg (54.2 lbs)
Net weight packaged With front protection glass: 40.1 kg (88.4 lbs)
Without front protection glass : 31.1 kg (68.6 lbs)
Mounting standard VESA (200 mm, 600 mm)
Certifications • ANSI/AAMI ES 60601-1:2005/(R)2012 and A1:2012, C1:2009/(R)2012
and A2:2010/(R)2012 - Med. El. Equip., Part 1: general req. for basic
safety and essential performance
• CAN/CSA-C22.2 No. 60601-1: 14 Medical Electrical Equipment - Part 1:
General Requirements for Basic Safety and Essential Performance
(Harmonized with Ed. 3.1)
• IEC 60601-1: 2012 Edition 3.1 (Medical electrical equipment – Part 1:
General requirements for basic safety and essential performance)
• EN 60601-1: 2006 + A1:2013 (Medical electrical equipment - Part 1:
General requirements for safety)
Important information