R5914634 /00 MDSC-852748
Immunity test
IEC 60601 test levels
Compliance level
Electromagnetic
environment – guidance
Voltage interruptions: 0%
residual voltage for 250
periods at 0°
Voltage interruptions: 0%
residual voltage for 250
periods at 0°
MDSC-8527 be powered
from an uninterruptible
power supply or a battery
Power frequency (50/60
Hz) magnetic field
IEC 61000-4-8
30 A/m 30 A/m Power frequency
magnetic fields should be
at levels characteristic of a
typical location in a typical
commercial or hospital
environment
Conducted RF
IEC 61000-4-6
3 Vrms (6 Vrms in ISM
bands)
150 kHz to 80 MHz
3 Vrms (6 Vrms in ISM
bands)
-
Radiated RF
IEC 61000-4-3
3 V/m
80 MHz to 2.7 GHz
3 V/m
Immunity to RF wireless communications equipment
Test
frequency
(MHz)
Band (MHz)
Service Modulation
Maximum
power (W)
Distance
(m)
Immunity
test level (V/
m)
385 380 – 390 TETRA 400 Pulse
modulation
18 Hz
1.8 0.3 27
450 430 – 470 GMRS 460,
FRS 460
FM ± 5 kHz
deviation
1 kHz sine
2 0.3 28
710 704 – 787 LTE Band
13, 17
Pulse
modulation
217 Hz
0.2 0.3 9
745
780
810 800 – 960 GSM 800/
900, TETRA
800, iDEN
820, CDMA
850, LTE
Band 5
Pulse
modulation
18 Hz
2 0.3 28
870
930
1720 1700 – 1990 GSM 1800,
CDMA 1900,
GSM 1900,
DECT, LTE
Band 1/3/4/
25, UMTS
Pulse
modulation
217 Hz
2 0.3 28
1845
1970
2450 2400 – 2570 Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation
217 Hz
2 0.3 28
5240 5100 – 5800 W LAN
802.11 a/n
Pulse
modulation
217 Hz
0.2 0.3 9
5500
5785
Important information