Instructions For Use
Warnings (continued)
Warning: Do not attempt to sterilize the Site~Rite* 6 Ultrasound System scanner or probe with
ethyleneoxideorheat sterilization methods. Damage to the equipment may occur.
Warning: Always properly dispose of dead battery packs in accordance with local regulations.
Improper disposal may present an environmental hazard.
Warning: Only qualified personnel should attempt to service this equipment. The Site~Rite*
6 Ultrasound System contains static sensitive components and circuits. Failure to
observe proper static control procedures may result in damage to the system.
Warning: The following actions void the warranty of the Site~Rite* 6 Ultrasound System and
may result in injury or equipment damage:
• Openingorservicingthescannerortheprobehousingbyanyoneotherthan
Bard Access Systems authorized service personnel.
• RemovalofsystemlabelsbyanyoneotherthanBardAccessSystemsauthorized
service personnel.
• OpeningorservicingthebasepowerunitortheA/Cadapterbyanyoneotherthan
Bard Access Systems authorized service personnel.
• ConnectingtheSite~Rite*6UltrasoundSystemscannertoanypowersourceother
than the Site~Rite* 6 Ultrasound System A/C adapter and/or base power unit.
• ConnectingtheSite~Rite*6UltrasoundSystemtoanyunauthorizedaccessory.
Refer to Section 1.2.
Warning: Inspect power cords for damage. If any of the prongs are damaged, use battery power
until a replacement cord is obtained.
Warning:
Verify that all accessories attached to the system comply to 60601 safety standards.
Non-compliance may result in increased patient risk.
Warning: Use only IEC or ISO approved safety devices outside the patient environment. Failure
to do so may damage the equipment.
Warning: Equipment that only relies on basic insulation should not be used with this system.
Failure to comply could result in increased patient risk.