9 Calibrating the Rechargeable Batteries
The Site~Rite* 6 Ultrasound System batteries may occasionally require calibration to ensure
the battery power meter is accurate. The following icon indicates that the Site~Rite* 6
Ultrasound System rechargeable batteries require calibration.
To calibrate the Site~Rite* 6 Ultrasound System Rechargeable Batteries:
1. Disconnect the Site~Rite* 6 Ultrasound System from A/C power.
2. Power on the Site~Rite* 6 Ultrasound System scanner and operate on battery power until the
system powers off.
3.ConnecttheSite~Rite*6UltrasoundSystemtoA/Cpowertorechargethebatteries.
Note: At least five hours of charge time is recommended to fully charge the Site~Rite* 6 Ultrasound System
batteries.
4. Disconnect the Site~Rite* 6 Ultrasound System from A/C power.
5. Power on the Site~Rite* 6 Ultrasound System and operate on battery power until the system
powers off.
Note: The batteries are now calibrated.
6. Connect the Site~Rite* 6 Ultrasound System to A/C power to recharge the batteries and continue
normal use.