How to fix Calibration Error on Bard Medical Equipment?
- AAndrew StanleySep 14, 2025
If you are experiencing a calibration error with your Bard Medical Equipment: 1. Restart the system. 2. Ensure the probe is held in the air during calibration.
How to fix Calibration Error on Bard Medical Equipment?
If you are experiencing a calibration error with your Bard Medical Equipment: 1. Restart the system. 2. Ensure the probe is held in the air during calibration.
How to fix Error Code EC:201-203 EC:205-206 EC:211 on Bard Site-Rite 8?
If you encounter Error Code EC:201-203, EC:205-206, or EC:211 on your Bard Medical Equipment, this indicates an ultrasound hardware failure. Try rebooting the system. If the error continues, further assistance is needed.
Why does my Bard Site-Rite 8 Medical Equipment show Error Code EC:101?
To resolve Error Code EC:101 on your Bard Medical Equipment, which indicates the system temperature is too high, shut down the system and allow it to cool down.
What does Error Code EC:110 mean on Bard Site-Rite 8?
If your Bard Medical Equipment displays Error Code EC:110, it means there's a battery error. Connect the AC adapter for operation and make sure the battery switch is in the correct position.
What to do if Cue™ Needle Tacking System Error appears on Bard Medical Equipment?
If you encounter a Cue™ Needle Tacking System Error on your Bard Medical Equipment, restart the system.
What to do for Error Code EC:403 on Bard Site-Rite 8?
If you encounter Error Code EC:403 on your Bard Medical Equipment, which indicates a Pinpoint™ GT Needle Technology Probe malfunction, turn off and restart the system.
What does Error Code EC:106 mean on Bard Site-Rite 8?
If you're experiencing Error Code EC:106 on your Bard Medical Equipment: 1. Ensure the device is connected to the desired network via WiFi or Ethernet. 2. If using a Static IP Address, verify the Static Network Settings. 3. Contact your network administrator.
How to resolve Error Code EC:107 on Bard Medical Equipment?
If you are seeing Error Code EC:107 on your Bard Medical Equipment: 1. Ensure you are connected to the desired network. 2. Ensure that the system DICOM settings are compatible with the destination PACS system settings. 3. Contact your network administrator.
What to do if Bard Medical Equipment system becomes unresponsive?
If your Bard Medical Equipment system becomes unresponsive, press and hold the power switch on the side of the device until the system shuts down (this will take approximately 10 seconds). Then, turn the system back on.
How do I fix Error Code EC:102 on Bard Medical Equipment?
If you are getting Error Code EC:102 on your Bard Medical Equipment, it means there's a USB storage device error. The USB storage device may be full or not compatible. Try replacing the USB storage device.
Critical warnings and cautions for operating the Site-Rite® 8 Ultrasound System safely.
Specific warnings and cautions for Cue™ and Pinpoint™ GT needle guidance.
Setup, activation, and interface usage for the Cue™ Needle Tracking System.
Setup, calibration, and tracking procedures for Pinpoint™ GT Needle Technology.
Resolving common errors for the Site-Rite® 8 Ultrasound System.
Specific troubleshooting for Cue™ and Pinpoint™ GT tracking systems.
Critical warnings and cautions for operating the Site-Rite® 8 Ultrasound System safely.
Specific warnings and cautions for Cue™ and Pinpoint™ GT needle guidance.
Setup, activation, and interface usage for the Cue™ Needle Tracking System.
Setup, calibration, and tracking procedures for Pinpoint™ GT Needle Technology.
Resolving common errors for the Site-Rite® 8 Ultrasound System.
Specific troubleshooting for Cue™ and Pinpoint™ GT tracking systems.
| Display Type | LCD |
|---|---|
| Transducer Ports | 1 |
| Battery Type | Rechargeable Lithium-Ion |
| Portability | Portable |
| Manufacturer | Bard |
| Power Source | AC Adapter or Battery |
| Image Modes | B-mode, M-mode |
| Needle Guidance | Yes |
| Intended Use | Vascular access procedures |