8 Ultrasound System Page 69
22. ACOUSTIC INFORMATION
22.1. Acoustic Output Summary Table
Description
of Probe
Operating
Mode
I
spta.X
(X denotes
statistically
determined
maximum)
FDA I
spta.3
Published
Values
MI X (X denotes
statistically
determined
maximum)
FDA MI Published
Values
Linear 32mm
Probe
B 10.1 mW/cm
2
Peripheral Vessel
< 720 mW/cm
2
Cardiac
< 430 mW/cm
2
Fetal Imaging & Other**
< 94 mW/cm
2
0.649
Peripheral Vessel
< 1.9
Cardiac
< 1.9
Fetal Imaging & Other**
< 1.9
20mm Linear
CueNeedle
Tracking
System Probe
B 11.1 mW/cm
2
Peripheral Vessel
< 720 mW/cm
2
Cardiac
< 430 mW/cm
2
Fetal Imaging & Other**
< 94 mW/cm
2
0.58
Peripheral Vessel
< 1.9
Cardiac
< 1.9
Fetal Imaging & Other**
< 1.9
Pinpoint™ GT
20mm Linear
Probe
B 22.0 mW/cm
2
Peripheral Vessel
< 720 mW/cm
2
Cardiac
< 430 mW/cm
2
Fetal Imaging & Other**
< 94 mW/cm
2
0.871
Peripheral Vessel
< 1.9
Cardiac
< 1.9
Fetal Imaging & Other**
< 1.9
** Abdominal, Intraoperative, Pediatric, Small Organ (breast, thyroid, testes, etc.), Neonate Cephalic, Adult Cephalic.