EasyManua.ls Logo

Bard Site-Rite 8 - Calibration

Bard Site-Rite 8
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Site~Rite
®
8 Ultrasound System Page 42
11.2.1. Calibration
Calibration may be required when the Performance Indicator (see Section 9.3.2) measures a poor magnetic
environment. To determine if calibration is necessary, hold the CueProbe Stationary in the air away from all
metal, magnetic, and electronic devices. If the Performance Indicator improves, calibration is not required.
For tips on how to improve the Performance Indicator when calibration is not required, press the
‘Performance Indicator Help button.
The following steps will calibrate the probe:
1. Select the calibrate button on the CueNeedle Tracking System settings screen.
2. Follow the on-screen instructions to hold the probe in the air, at least 3 feet from any metal, magnetic, or
electronic devices (including watches, cell phones, etc.).
3. While holding the probe still in the air, press the calibration butt. Maintain this position until the system
indicates that calibration is complete.
4. Upon completion of calibration, the Performance Indicator should improve. A Performance Indicator of 3 bars
or higher is recommended before beginning any procedure.
Note: Maintain a distance of 4 inches between the Cue™ probe and Cue™ Activator to minimize the
potential need for calibration.

Table of Contents

Related product manuals