Site~Rite
®
8 Ultrasound System Page 71
22.3. Acoustic Output Reporting Table – Linear 20mm Cue™ Needle Tracking System Probe
Transducer Model: Linear 20mm Cue™ Needle Tracking System Probe
Index Label MI
TIC
Scan
Non-scan
Non-scan
Maximum index value
0.517 0.134 - - - (a)
Associated
acoustic
parameter
r.3
0
min of [W.
3
(z
1
), I
TA,3
(z
1
)]
1
bp
-
sp
1.40 -
eq
sp
c
Dim of A
aprt
7420
r
max
eq
max
-
Focal Length
x
0.107 - -
pa.3
Operating
Control
Conditions
1.92cm scan length
64 lines per scan
Focal depth 20mm
Frequency 7.5 MHz, 10 MHz
Note 1: Information need not be provided for any formulation of TIS not yielding the maximum value of TIS for that mode.
Note 2: Information need not be provided regarding TIC for any TRANSDUCER ASSEMBLY not intended for transcranial or neonatal
cephalic uses.
Note 3: Information on MI and TI need not be provided if the equipment is not capable of exceeding a TI or MI of 1.0.
(a) Intended use does not include cephalic so TIC is not computed
# No data reported.