11
Instructions For Use
Caution: Only apply commercially available ultrasonic couplant, which has been specifically
formulated for use in medical applications, to the acoustic window (or face) of the
probe.
Caution: Use water or rubbing alcohol and a soft cloth to remove couplant from the acoustic
window (or face) of the probe. Failure to do so may scratch the acoustic window.
Caution: Do not to allow ultrasonic couplant to dry on the acoustic window (or face) of the
probe. If the couplant should dry, use water or rubbing alcohol and a soft cloth to
remove it. Never use a tool of any kind to remove dry couplant from the acoustic
window (or face) of the probe.
Caution: Somecommerciallyavailableprobecoverscontainlatex.Naturalrubberlatexmay
cause allergic reactions. Refer to the US FDA alert titled: “Medical Alert: Allergic
ReactionstoLatex-ContainingMedicalDevices”,issuedMarch29,1991.
Bard Access Systems distributes sterile probe covers and needle guide kits that do
notcontainlatex.
Caution: Do not force the probe connector. Damage to the connector and system could result.
Caution: Always snap the needle guides on to the probe hook. Do not slide the needle guide
on to the needle guide hook, as the sterile sheath may tear.
Caution: Donotsubjecttheprobetoexcessivevibration.Vibrationmaydislodgesensitive
components and cause intermittent or unpredictable operation.
Caution: Prior to each use, inspect the integrity of all power cords and connectors as well as
the integrity of the unit itself. If any problems are found, discontinue use immediately
and contact an authorized service representative. Use of a damaged power cord could
damage the machine.
Caution: During use, the A/C connector needs to be easily accessible. In case of emergency
remove the power cord as soon as possible.
Caution: To avoid unnecessary strain, use the device in a comfortable manner.