For Main Firmware Version 1.90
Calibrating and Adjusting MC5 / 011912
Calibration and
Adjustment Instructions
for MC5 Multifunction Calibrator
Enable Adjustment Mode
Select Quantity and Function/Port
Calibrate (Compare against a
more accurate instrument)
Adjustment required?
Store adjustment data
Enter reference values to
adjust a Quantity in the calibrator
More Quantities to calibrate?
Did one or
more Quantities in
all modules
require
adjustment?
Enter the calibration date and
Calibration Laboratory data in
the Module Settings window
yes
no
no
yes
no
yes
Start
End
More Quantities to adjust?
no
yes
Question and Answer IconNeed help?

Do you have a question about the BEAMEX MC5 and is the answer not in the manual?

BEAMEX MC5 Specifications

General IconGeneral
TypeMultifunction Calibrator
Current MeasurementYes
Voltage MeasurementYes
CommunicationYes
Pressure ModulesSupports external pressure modules
Temperature MeasurementRTD, TC
Electrical MeasurementFrequency
Output ParametersVoltage, current, frequency, pulse
Communication InterfaceHART, RS-232
Power SupplyRechargeable battery
DisplayGraphical LCD
Operating Temperature-10°C to +50°C
Storage Temperature-20°C to +60°C
Humidity10% to 90% RH (non-condensing)
Pressure MeasurementInternal and external pressure measurement

Summary

General Information

Equipment Requirements for Calibration

Lists the necessary equipment for performing the calibration procedure.

Calibrating and Adjusting the MC5 Calibrator

Enabling the Adjustment Mode

Steps to enable the adjustment mode for calibrator adjustments.

Adjusting Measurement Ranges

Procedures for adjusting the calibrator's measurement ranges for various modules.

Adjusting Generation and Simulation Ranges

Procedures for adjusting the calibrator's generation and simulation ranges.

Saving Calibration Adjustment Data

Instructions on how to save the calibration adjustment data.

Setting Calibration Date Without Adjustment

Procedure to set calibration date when no adjustments were needed.

Ending Adjustment Mode

Steps to exit adjustment mode and return to normal operation.