FBXXX___tgeng0000
Field Temperature Block series
Beamex
®
FB150, FB350, FB660
User Guide

Table of Contents

Questions and Answers

  • E
    Eric NixonAug 12, 2025
    How to fix BEAMEX Test Equipment that does not power up?
    • T
      Tammy KiddAug 13, 2025
      If your BEAMEX Test Equipment does not power up, there might be a power surge or a component failure. First, check the fuses and replace any blown fuses, but do not use a fuse with a higher current rating. Also, verify that the power cord is properly plugged into the instrument and that the power circuit is active. Check if the circuit breaker has tripped and press the reset button to reset it.
  • C
    Caleb HendersonAug 17, 2025
    What to do if the display shows cutout on BEAMEX FB150 Test Equipment?
    • Q
      quinnvictorAug 17, 2025
      If the display of your BEAMEX Test Equipment shows cutout, it could be because the temperature block has exceeded the temperature set in the soft cutout menu or the maximum operating temperature. To resolve this, reset the instrument using the SET PT. key to clear the cutout mode and reactivate the instrument. You may also need to check and adjust the cutout setting by navigating to MENU|TEMPSETUP|CUTOUT.
  • C
    christopherwhiteAug 20, 2025
    Why is the temperature readout incorrect on my BEAMEX FB150?
    • D
      Devin ClarkAug 20, 2025
      If the temperature readout on your BEAMEX Test Equipment is not the actual temperature, ensure that all operating parameters for the temperature block, reference thermometer, and probe parameters match the Report of Certification. Also, check for sources of electrical interference, such as motors or welding equipment, and try shielding wires or changing the location to minimize interference.
  • C
    Christopher CarterAug 25, 2025
    How to fix a blank display on BEAMEX Test Equipment that powers up?
    • K
      Katie KeithAug 25, 2025
      If the display on your BEAMEX Test Equipment is blank but the instrument powers up, try adjusting the screen contrast using the down arrow key to lighten the screen.
  • J
    johngriffinAug 27, 2025
    What to do if probes are stuck in the well at low temperatures in BEAMEX (FB150)?
    • J
      Jennifer LawrenceAug 28, 2025
      If you are experiencing probes stuck in the well of your FB150 BEAMEX Test Equipment at low temperatures, it is likely due to moisture. To remove the probes, set the temperature high enough to melt the ice. Then, set the set-point to +100°C to allow the moisture to evaporate out of the system.
  • K
    Kathy BrockAug 31, 2025
    How to remove an insert stuck in well of BEAMEX (FB150)?
    • K
      Kathleen PerezAug 31, 2025
      If the insert is stuck in the well of your FB150 BEAMEX Test Equipment, it may be due to hard water build-up. To remove it, keep the unit in a cold environment (less than 21°C) and set the unit to 100°C. While it's heating, try pulling on the insert when the temperature reaches approximately 50°C to 70°C.
  • A
    Adam RoblesSep 4, 2025
    Why does my BEAMEX Test Equipment heat slowly?
    • P
      petersonjeremySep 4, 2025
      If your BEAMEX Test Equipment heats slowly, check the Scan Rate settings. The scan rate might be set too low for your current application.
  • D
    David WatsonSep 8, 2025
    What causes an abnormal temperature or '...' on the Ref Probe of BEAMEX FB150 Test Equipment?
    • L
      langdeannaSep 8, 2025
      Check the Probe type setting in the Reference Probe Setup menu to ensure it is correct. Also, check all associated parameters and ensure that the 4 probe wires are correctly connected and not shorted inside the connector.