FBXXX Field Temperature Block
Non-SCPI Process Commands
66
SYST:TIME[?] [<hh,mm,ss] (-R model only)
NOTE: This command is conditionally protected and requires a password to set
it.
Read or set the System Time <hh,mm,ss> (24 hr time only)
Range: hh = {0 – 23}
mm = {0 – 59}
ss = {0 – 59}
Default: Current Time – Pietarsaari, Finland
Read Example: SYST:TIME?
Response: 23,51,05
Set Example: SYST:TIME 14,15,05
UNIT:TEMP[?] [n]
Read or set the display temperature units, where “n” is a character “C” or “F”. Default:
C
Read Example: UNIT:TEMP?
Response: C
Depending on instruments setting, a C (Celsius) or F (Fahrenheit) is returned.
Set Example: UNIT:TEMP F
6.5 Non-SCPI Process Commands
This section contains Non-SCPI commands. These are available for users that require
Non-SCPI commands for their application. These commands are used differently
from the SCPI commands discussed in the previous section, the command protocol
and response is different. These commands do not require a question mark (?) for a
query, and respond to a query by rst outputting the command and colon before the
data. These commands are not password protected. The associated SCPI command is
referenced where appropriate.
6.6 Non-SCPI Commands by Function or Group
SCREEN
PARAMETER COMMAND
PASSWORD
PROTECTION READ/WRITE
Setup - Communication DUPLEX du None R
LINEFEED lf None R/W
SAMPLE RATE sa None R/W