65
Digital communication interface
Serial Commands - Alphabetic Listing
This command has no response.
This command disables the system password protection.
SYST:PASS:CEN [n]
Enable access to password protected setting commands, where “n” is a four digit pass-
word. Range = {0000 – 9999}; Default: 1234
Example: SYST:PASS:CEN 1234
This command has no response.
This command enables the system password. This password needs to be enabled in
order to use the conditionally protected commands. When the power of the instrument
is cycled, system password protection is disabled.
SYST:PASS:CEN:STAT?
Read the access state of password protected setting commands.
Example: SYST:PASS:CEN:STAT?
Response: 0
This command reports the current status of the system password.
SYST:PASS:NEW <n>|DEF
NOTE: This command is unconditionally protected, which requires a password
to set it.
Set the password, where “n” is the new four digit password. Range = {0000 – 9999};
Default: 1234
Example: SYST:PASS:NEW 1234
This command has no response.
This command allows the user to set the system password.
SYST:PASS:PROT[?] [0|1]
Read or set password protection level., [0] = low, [1] = high
Read Example: SYST:PASS:PROT?
Response: 0
Set Example: SYST:PASS:PROT 1