FBXXX Field Temperature Block
Serial Commands - Alphabetic Listing
64
SYST:KLOC[?] [n]
NOTE: This command is unconditionally protected and requires a password to
set it.
Read or set the keypad lockout; [0] = unlock, and [1] = lock. Default: 0 (Unlock)
Read Example: SYST:KLOCK?
Response: 1
Set Example: SYST:KLOC 1
This command locks or unlocks the system keypad providing control only through the
serial interface (RS-232 port) or the keypad.
SYST:LANG <lang>
Set the display language. The available languages are dependent upon which version
of the product is supplied. The version is dependent upon the nal destination and
conguration.
European: ENGLish (default), FRENch, SPANish, ITALian, GERMan
Russian: RUSSian (default), ENGLish
Asian: ENGLis (default), CHINese, JAPanese
Example: SYST:LANG SPAN
SYST:LANG:CAT?
Read the available display languages. The available languages are dependent upon
which version of the product is supplied. The version is dependent upon the nal desti-
nation and conguration.
European: ENGLish (default), FRENch, SPANish, ITALian, GERMan
Russian: RUSSian (default), ENGLish
Asian: ENGLish (default), CHINese, JAPanese
Example European: SYST:LANG:CAT?
Response: “ENGL”,”FREN”,”SPAN”,”ITAL”,”GERM”
Example Russian: SYST:LANG:CAT?
Response: “RUSS”,”ENGL”
Example Asian: SYST:LANG:CAT?
Response: “ENGL”,”JAP”,”CHIN”
SYST:PASS:CDIS
Disable access to password protected setting commands
Example: SYST:PASS:CDIS