FBXXX Field Temperature Blocks
Main Menu
30
5.2.1.2.2 HARD CUTOUT
The HARD CUTOUT parameter is a view only function and indicates the factory set-
ting for the hard cutout. The Hard Cutout is not user settable.
5.2.2 Prog Menu
The PROG MENU (PROGRAM MENU) allows access to the automated and manual
program selections.
5.2.2.1 RUN PROG
The RUN PROG (RUN PROGRAM) allows the user to access program status
features.
5.2.2.1.1 TEST STATUS
The TEST STATUS option controls the state of the program. The user selects to Run
the program or to turn the program Off.
5.2.2.1.2 RECORD DATA
The RECORD DATA option allows the user to select to record the data from the test
(Yes) or not to record data (No).
5.2.2.1.3 TEST ID
The TEST ID (Identication) allows the user to enter a Test ID number for the current
test. The Test ID can be an alpha numeric entry up to 16 characters in length.
5.2.2.2 RAMP/SOAK
The RAMP/SOAK feature automatically cycles the Field Temperature Block between
temperatures while holding at each temperature for the length of time set by the user.
5.2.2.2.1 NO. SETPOINTS
The NO. SETPOINTS is the number of set-points dened for a given program. The
number of set-points for each program can be set from 1 to 8 and vary depending on
the needs of the user. Set the maximum number of set-points needed for the program
selected. Once the number of set-points is selected, press “ENTER” to accept the new
setting.
5.2.2.2.2 SOAK TIME
The SOAK TIME parameter is the number of minutes that each of the program set-
points is maintained. The time starts when the temperature settles to within the speci-
ed stability. The stability limit is set in the TEMP SETUP|SETUP|STABLE LIMIT
window.