23
Menu Structure
Temp Setup Menu
4 Menu Structure
4.1 Temp Setup Menu
Figure 10
FB150/ FB350/ FB660 MAIN MENU
F1 F2 F3 F4
TEMP SETUP PROG MENU SYSTEM MENU INPUT SETUP (-R ONLY)
F1 Setup
Scan Rate: <edit> Degrees per minute at which the Block increases or decreases temperature
Stability Limit: <edit> The defined minimum stability over a 2 minute period before alarm is activated
Stability Alarm: (ON/OFF) <select> Audible alarm indicating Block is stable
F2 View Cutout
Soft Cutout: <edit> User adjustable safety cutout
Hard Cutout: <View Only> Factory set safety cutout
Hot Keys (while viewing main screen)
SETPoinT- Set Point Key
Setpoint: <Edit> Set point temperature
ENTER – <Enable control of the unit>
F1 – SELECT PRESET <1-8> <select>
F1 – EDIT PRESETS <1-8> <edit>
F4 – SAVE/DISABLE System <disables control of instrument>
ºC / ºF Key - Units: <°C, °F> Changes temperature units
Up/ Down Arrow Keys <toggle> <adjust contrast>
Up Key: Darker
Down Key: Lighter
F1 & F4 Keys (Same Time)
< Reset to English Language >
F1 & F3 Keys (Same Time)
< Turn off key press beep >
Code Update Mode Keys
ENTER & EXIT Keys (hold during power up) <initiate code update mode> Allows instrument software to be updated
Main Menu - Temp SetUp