31
Controller operation
Main Menu
5.2.2.2.3 NO. CYCLES
The NO. CYCLES parameter is the number of times that the program is repeated.
5.2.2.2.4 DIRECTION
The DIRECTION parameter controls whether the set-points are sequenced in one
direction, 1-8, or both directions, 1-8 and 8-1, before the sequence is repeated. If the
both directions option is selected, the program sequences from the rst set-point to the
last and then reverses direction sequencing from the last to the rst.
5.2.2.2.5 PASS TOLERANCE
The PASS TOLERANCE is the allowable tolerance condition for the test and is used
to highlight test points that have large errors.
5.2.2.2.6 SETPOINTS
The SETPOINTS menu allows the user to set each of the set-points for the program.
Only the number of set-points dened by NO SETPOINTS will be displayed. Set-
points can be quickly selected using the Up/Down arrow keys to scroll through the set-
points. Press “Enter” to activate the set-point and make it editable. Once editable, use
the Up/Down arrow keys to enter the values and the Left/Right arrow keys to scroll
through the digits in the value. Press “Enter” to accept the value entered.
5.2.2.2.6.1 SETPOINT 1 – SETPOINT 8 (Depending on NO. SETPOINTS defined)
The SETPOINT n parameter is the designated temperatures for the set-points selected
for the program.
5.2.2.3 TEST RESULTS
The TEST RESULTS MENU allows the user access to the tests parameters.
5.2.2.3.1 VIEW TESTS
The VIEW TESTS menu allows the user to view the results of up to 16 tests.
5.2.2.3.1.1 TEST ID
The TEST ID (TEST IDENTIFICATION) parameter allows the user to select from 16
tests to view.
5.2.2.3.1.2 TYPE
The TYPE parameter allows the user to select between viewing Ramp & Soak results.
5.2.2.3.1.3 DATE (VIEW ONLY)
The DATE the selected test was performed.