63
Digital communication interface
Serial Commands - Alphabetic Listing
Response: 0
Set Example: SYST:COMM:SER:LIN 1
This command enables or disables line feed.
SYST:CONF:MOD?
Read the presence of the -R module; [0] if no -R sensor module, [1] if -R sensor card
is installed
Example: SYST:CONF:MOD?
Response: 1
If the -R sensor module is installed, the instrument is a -R Model.
SYST:DATE[?] [<year>,<month>,<day>] (-R model only)
NOTE: This command is conditionally protected and requires a password to set
it.
Read or set the System Date Setting using numbers separated by commas
(yyyy,mm,dd). Default: <Blank>
Read Example: SYST:DATE?
Response: 2007,05,24
Set Example: SYST:DATE 2007,05,24
SYST:DEC:FORM[?] [n]
Read or set the decimal format, where “n” is period [0], comma [1]. Default: 0
(Period)
Read Example: SYST:DEC:FORM?
Response: 0
Set Example: SYST:DEC:FORM 1
SYST:ERR?
Read the most resent error from the error queue
Example: SYST:ERR?
Response: command protected
This command response reports the errors in the error queue.