FBXXX Field Temperature Block
Serial Commands - Alphabetic Listing
62
The system beeper should make an audible sound in response to this command.
SYST:BEEP:KEYB[?] [n]
Read or set the keyboard beep function, 0=Off, 1=On. Default: 1
Read Example: SYST:BEEP:KEYB?
Response: 1
Set Example: SYST:BEEP:KEYB 1
Turns the keyboard beep function on or off.
SYST:CODE:LANG?
Read the language set option: 1: European; 2: Russian; 3: Asian. The available lan-
guages are dependent upon which version of the product is supplied. The version is
dependent upon the nal destination and conguration.
European: ENGLish (default), FRENch, SPANish, ITALian, GERMan
Russian: RUSSian (default), ENGLish
Asian: ENGLish (default), CHINese, JAPanese
Example: SYST:CODE:LANG?
Response: 3
SYST:CODE:VERS?
Read the main code version
Example: SYST:CODE:VERS?
Response: 1.10
Provides the user with the version of the main processor code.
SYST:COMM:SER:BAUD[?] [<baud>]
Read or set serial interface baud rate where “baud” is a standard baud rate value.
Range baud = {1200, 2400, 4800, 9600, 19200, and 38400}; Default: 9600
Read Example: SYST:COMM:SER:BAUD?
Response: 2400
Set Example: SYST:COMM:SER:BAUD 9600
SYST:COMM:SER:LIN[?] [n]
Set serial interface linefeed enable, where “n” is a value 1 or 0. [0] = LF OFF, [1] = LF
ON; Default: 1 (OFF)
Read Example: SYST:COMM:SER:LIN?