Digital communication interface
Serial Commands - Alphabetic Listing
Set Example: SOUR:SPO 100.00
This command reads or sets the value of the control set point based on the system
temperature units.
SOUR:STAB:BEEP[?] [n]
Read or set the stability alert (beep) enable where “n” is a value 0 or 1. [0] is disable,
[1] is enable beep. Default:1 (Enable Beep)
Read Example: SOUR:STAB:BEEP?
Response: 1
Set Example: SOUR:STAB:BEEP 0
Enable or disable the audible stability alert.
SOUR:STAB:DAT?
Read the control temperature stability, °C or °F
Example: SOUR:STAB:DAT?
Response: 0.306
The controller stability is returned.
SOUR:STAB:LIM[?] [n]
Read or set the control temperature stability limit, °C or °F where “n” is a positive real
value. Range = {0.01 to 9.99 (°C)}; Default: 0.05 (°C)
Read Example: SOUR:STAB:LIM?
Response: 0.05
Set Example: SOUR:STAB:LIM 0.03
Read or set the control stability limit.
SOUR:STAB:TEST?
Read the temperature stability test results. Stable = 1; Unstable = 0
Example: SOUR:STAB:TEST?
Response: 0
A value of 0 is returned if the controller is not stable at the current set-point. Otherwise
a value of 1 is returned if the controller is stable at the current set-point.
SYST:BEEP:IMM
Beep the system beeper
Example: SYST:BEEP:IMM