FBXXX Field Temperature Block
Serial Commands - Alphabetic Listing
60
Read Example: SOUR:SENS:CAL:PAR1?
Response: 0.0
Set Example: SOUR:SENS:CAL:PAR2 0.02
This command reads or sets the calibration parameter value for main control.
SOUR:SENS:CAL:TEMP<x>?
Read the required calibration temperature (°C) corresponding to a calibration parame-
ter where “x” is a numeric value indicating the parameter [1] = TEMP1, [2] = TEMP2,
AND [3] = TEMP3. Range = {1-3}; Default = 1
Example: SOUR:SENS:CAL:TEMP1?
Response: 40
SOUR:SENS:DATA? [TEMP]
Read the control temperature, °C or °F
Example: SOUR:SENS:DATA? or SOUR:SENS:DATA? TEMP
Response: 30.285°C (current control temp)
The current control temperature is returned if the above or if TEMP is appended to the
end of the example.
SOUR:SENS:DATA? [RES]
Read the control sensor resistance
Example: SOUR:SENS:DATA? RES
Response: 111.28
When RES is appended to the end of the example above, the internal sensor resistance
is returned.
SOUR:SPO[?] [n]
Set the control set-point, °C or °F, where “n” is a real value with acceptance limits
based on the model
Parameter Min Max Default
FB150 –25.00 150.00 25.00
FB350 25.00 350.00 25.00
FB660 25.00 660.00 25.00
Read Example: SOUR:SPO?
Response: 50.000