EasyManua.ls Logo

BEAMEX FB150 - Page 66

Default Icon
78 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
FBXXX Field Temperature Block
Serial Commands - Alphabetic Listing
60
Read Example: SOUR:SENS:CAL:PAR1?
Response: 0.0
Set Example: SOUR:SENS:CAL:PAR2 0.02
This command reads or sets the calibration parameter value for main control.
SOUR:SENS:CAL:TEMP<x>?
Read the required calibration temperature (°C) corresponding to a calibration parame-
ter where “x” is a numeric value indicating the parameter [1] = TEMP1, [2] = TEMP2,
AND [3] = TEMP3. Range = {1-3}; Default = 1
Example: SOUR:SENS:CAL:TEMP1?
Response: 40
SOUR:SENS:DATA? [TEMP]
Read the control temperature, °C or °F
Example: SOUR:SENS:DATA? or SOUR:SENS:DATA? TEMP
Response: 30.285°C (current control temp)
The current control temperature is returned if the above or if TEMP is appended to the
end of the example.
SOUR:SENS:DATA? [RES]
Read the control sensor resistance
Example: SOUR:SENS:DATA? RES
Response: 111.28
When RES is appended to the end of the example above, the internal sensor resistance
is returned.
SOUR:SPO[?] [n]
Set the control set-point, °C or °F, where “n” is a real value with acceptance limits
based on the model
Parameter Min Max Default
FB150 –25.00 150.00 25.00
FB350 25.00 350.00 25.00
FB660 25.00 660.00 25.00
Read Example: SOUR:SPO?
Response: 50.000

Table of Contents