Digital communication interface
Non-SCPI Commands by Function or Group
Temperature Settings HIGH LIMIT hl None R
SET POINT s None R/W
TEMPERATURE t None R
System Information VERSION *ver None R
System Setup °C/°F u None R/W
*ver
Read the Model number and Main code version (Model Number, Firmware version). A
question mark (?) is not required to query this command.
Example: *ver
ver. FB150, 1.00
du
Read or set serial interface echo enable, on (1) or off (0).
The FB150, FB350 and FB660 do not support the Full duplex mode. The response will
be the command string and a colon followed by “Half”.
Read Example: du
du: HALF
Set Example: du 1
This command enables or disables the echo.
hl
Read the maximum temperature setting for the unit. This command is query only and
responds with the command string and a Colon followed by the maximum temperature
and associated units.
Read Example: hl
hl: 660.00 C
lf [n]
Read or set the serial interface linefeed enable, where “n” is a value 1 or 0. [0] = LF
OFF, [1] = LF ON. The default setting is Off. (Off and on may be used in place of 0
and 1 respectively). If “n” is left blank, the command will be treated as a query. This
query responds with the command string and a Colon followed by the LF setting.
Refer to SYST:COMM:SER:LIN command.
Read Example: lf
lf: OFF
Set Example: lf on